TOP >Product Information

PEMGRAPH

Features


Whole Breast in the Field of View

    The detector consists of two 20-cm-long parallel plates (20×14 cm2 each) and is designed so that the detector assembly can be moved vertically in 25-mm increments.
    The detector spacing can be optimized to the size of the breast, allowing the entire breast, including the axillary and chest-wall regions, to be
    included in the field of view with no blind areas.
Reduced Anxiety During Examination

    Unlike X-ray mammography, PEMGRAPH does not require strong breast compression, and examinations can also be performed while the patient is wearing an examination gown.
    This helps substantially reduce anxiety and concern associated with breast cancer examinations.
Improved Spatial Resolution

    The detector uses Pr:LuAG (Pr:Lu3Al5O12) scintillators with excellent timing resolution. Compared with whole-body PET systems,
    PEMGRAPH has a smaller detector separation, and the use of fast scintillators helps improve imaging performance.
    Furthermore, by optimizing the Pr:LuAG scintillator crystal size for the detector configuration,
    a spatial resolution of 2.0 mm or better has been achieved.
Compact Design

    The main unit has a compact installation footprint of 50×165 cm2, approximately 50% of the area of a standard tatami mat (91×182 cm2).
    The detector positioning mechanism is motorless, helping reduce the overall weight.
    Vertical detector movement and ±90-degree rotation make breast positioning easy.
Time Saving

    Each breast can be scanned in approximately 3–5 minutes, and both breasts can be examined in about 10 minutes in a typical protocol including positioning.
    Image reconstruction takes approximately 1 minute, enabling images to be reviewed immediately.

Specifications


Generic Name Positron CT System for Nuclear Medicine Diagnosis (40644000)
Product Name Breast PET PEMGRAPH
Medical Device Approval No. 22600BZX00513000
Electrical Rating Supply Voltage AC100V 50/60Hz
Power Input 600 VA (at maximum load)
Equipment Classification Type of Protection Class I equipment
Degree of Protection Equipment with a Type B applied part
Performance Spatial Resolution FWHM 2.0㎜ or less
FWTM 3.5㎜ or less
Peak Count Rate Peak true coincidence count rate: 50 kcps or more
Peak noise-equivalent count rate: 30 kcps or more
Sensitivity 5.0cps/kBq
Attenuation Correction Accuracy Contrast Target Size 4× Background
φ 3.7㎜ hot sphere 10% or more
φ 4.8㎜ hot sphere 25% or more
φ 6.0㎜ hot sphere 40% or more
φ 6.0㎜ cold sphere 50% or more
Fixation with Plate Maximum pressing force of fixation plate: 55 N ±10%
Main Unit Dimensions W 502㎜ × D 1635㎜ × H (max.) 1655㎜
Main Unit Mass 350kg
Configuration Main Unit PEMGRAPH main unit
Isolation transformer    Foot switch (IPX7)
Console Data-acquisition PC
Image-reconstruction PC
Operation PC
One monitor for the above PCs, keyboard, and mouse
Accessories Fixation plate

Back to top

Detector Unit Configuration


Detector

Four detector units are incorporated into each of the upper and lower detector assemblies.

Properties of Pr:LuAG

Back to top

Detector Movement Range


Detector Rotation Range


Vertical Detector Movement Range


Detector Spacing Adjustment Range


Back to top

Imaging Positions


CC position

Pr:LuAG

MLO position

Pr:LuAG

Back to top

Layout



System Dimensions

Weight: 350 kg
Dimensions: W 502 mm × H 1655 mm × D 1635 mm

Isolation Transformer
Weight: 16 kg
Dimensions: W 220 mm × H 182 mm ×
D 355 mm Tolerance


Back to top

PEM Imaging Examples

Clinical Evaluation of Spatial Resolution [1]
Clinical Evaluation of Spatial Resolution [2]

Sendai Imaging Clinic
— clinical imaging and evaluation

Back to top

Small-Animal PET

Mirai-PET

World’s first implementation of a CLS (Crosshair Light-Sharing) detector*

Simultaneous DOI and TOF capability

Spatial resolution of less than 1 mm
pdf   pdf                                                                
* The CLS detector is a technology put into practical use by the National Institutes for Quantum Science and Technology (QST).

Back to top

Radiation Measurement

Gamma Spotter II

Compact, lightweight, easy to carry, and simple to operate.
Reliable Gamma Spotter performance.
Three measurement modes are available.
pdf   pdf                                                                

Gamma Spotter F

A next-generation inspection system designed to further enhance food safety and provide thorough inspection.
pdf


Back to top

Scintillator Crystals

Ce:GAGG

Ce:GAGG is a scintillator crystal with high light output and high density, and it is non-hygroscopic.
Short-decay-time and enhanced-energy-resolution versions are also available
to suit your application. Please contact us for details.
pdf

Pr:LuAG

Pr:LuAG is a scintillator crystal with a short fluorescence decay time and excellent energy resolution.
It is also used in our Breast PET PEMGRAPH
system.
pdf